Table 6 Number and percentage of initial and final reports per category

From: Quality analysis of manufacturer's incident reports regarding medical devices

 

Number of EU-based initial reports

%

Number of non-EU-based initial reports

%

Number of EU-based final reports

%

Number of non-EU-based final reports

%

Excellent

98

21.03

45

40.18

262

57.21

73

67.59

Good

246

52.79

51

45.54

50

10.92

6

5.56

Medium

71

15.24

10

8.93

62

13.54

7

6.48

Qualified

32

6.87

4

3.57

12

2.62

1

0.93

Unqualified

19

4.08

2

1.79

72

15.72

21

19.44

  1. Median quality of the initial reports from manufacturers based on EU countries and third countries were 84.62% and 87.64%, respectively; the distributions in the two groups differ significantly (Mann–Whitney U = 38,598, n1 = 466, n2 = 112, p = 0.0001 two-tailed)
  2. Median quality of the final reports from manufacturers based on EU countries and third countries were 90.91% and 91.67%, respectively; the distributions in the two groups do not differ significantly (Mann–Whitney U = 30,191, n1 = 458, n2 = 108, p = 0.7746 two-tailed)